Back to Top

Hi, Guest!

  LOKASI :  Kota Bogor

Bergabung Selama : 17 Tahun

BAGIKAN :   

Bagikan :
Kontak Kami  
Instrumentation & Valves

MODEL 2001LC Trace Oxygen Gas Analyzer.

Update Terakhir
:
10 / 01 / 2024
Min. Pembelian
:
1 Unit
Dilihat Sebanyak
:
29 kali

Harga

CALL
Bagikan
:
*Instrumentation, Control System, Automation, Cooler & Heat Exchanger, Electrical, Mechanical, Analyzers, Fabricator & Engineering Service & Training*

Detail MODEL 2001LC Trace Oxygen Gas Analyzer.

Technical Specifications

Physical & Electrical Specifications
WEIGHT MOUNTING GAS CONNECTIONS WETTED
PARTS
5.0 lbs
(2.3 kg)
Panel Mount ¼” 316 S.S.
compression
fittings
316 S.S. fittings, electro-less nickel-plated cell block, gold-plated contacts, acrylic flow meter & O-rings (Viton, kalrez, and Buna-N)
POWER
DC AC
90-240VAC <10W
model-2001lc dimensions
Operational Specifications
OUTPUT RANGES
(USER SELECTABLE)
AMBIENT OPERATIONAL
TEMPERATURE
FLOW RATE
(RECOMMENDED)
INLET GAS
PRESSURE
0 – 10 ppm
0 – 50 ppm
0 – 100 ppm
0 – 500 ppm
0 – 1000 ppm
0 – 5000 ppm
0 – 1.0%
0 – 5.0%
0 – 10.0%
0 – 25.0%
41°F to 113°F
(5°C to 45°C)
0.1 to 5.0 SCFH
(0.05 – 2.4 Lpm)
0.5 − 150 psig
(0.03 − 6.9 bar)
Sensor & Analyzer Performance Specifications
T-2 SENSOR P-2 SENSOR P-3 SENSOR P-4 SENSOR 90% UPSCALE RESPONSE TIMES
FOR SPECIFIC RANGES
REPEATABILITY OPERATIONAL CONDITIONS
Max Range
0 –10,000 ppm

Special Conditions
<10 ppm H2S, 100% CO2
Max Range
0 – 50%
90% upscale response:
8 sec
Special Conditions:
O2 in inert gas
Max Range
0 – 25%
90% upscale response:
30 sec
Special Conditions:
Up to 100% CO2
Max Range
0 – 100%
90% upscale response:
30 sec
Special Conditions:
Enriched Oxygen
<25 sec for 0 – 10 ppm
<10 sec for 0 - 100 ppm
<10 sec for 0 - 1000 ppm
±1% of range or
±0.2 ppm of oxygen,
whichever is greater
Pollution degree 2, Installation category II
Analyzer Key Features
ELIMINATOR CELL BLOCK™ Contains a Complete Sample System with all components, including the flow meter, metering valve, pressure sensor and 3-way Selection (Sample/Span/Off) Valve, fully integrated
ALARMS
2 fully adjustable concentration alarms with configurable Alarm logic and Relay Contacts, featuring:
  • Alarm Bypass
ANALOG OUTPUTS 4 − 20mA isolated output signals

Features & Benefits

Eliminator Cell Block with Component Call-OutsTransparent View of ELIMINATOR CELL BLOCK™

ELIMINATOR CELL BLOCK™ FOR A LEAK-FREE DESIGN

The patented ELIMINATOR CELL BLOCK™ provides a unique, compact sample system approach and a flow-efficient sensor pocket machined into a solid metallic block. Precision cross-drilled intersecting gas passages eliminate the need for long lengths of tubing and leak-prone compression fittings. Additionally, we have engineered the metering valve and flow meter to be an integral part of the ELIMINATOR CELL BLOCK. It also provides the user with direct front panel access for installing and replacing sensors, and the ability to air calibrate the analyzer, all without the need for disassembly or tools.

Image of sensor rendering exploded to show individual elementsExploded View of Sensor

PROPRIETARY SENSOR TECHNOLOGY

The 2001LC combined with AMI’s patented oxygen sensors that deliver unmatched performance for ultra-fast response times, accuracy, and reliability. Every sensor undergoes an intense battery of tests to ensure they meet our strict quality control standards. Our trace and percent oxygen sensors operate safely, accurately, and reliably in virtually all gas streams.

Model 2001LC Front PanelModel 2001LC Front Panel

EXCEPTIONAL VALUE

The MODEL 2001LC is the perfect low-cost solution for trace O2 measurements requiring a complete sample system and limited key features. It offers the same reliability, accuracy and repeatability of other AMI Analyzers while including multiple output ranges, dual independent alarms, and intuitive controls. This virtually maintenance-free oxygen analyzer is a very cost-effective option for general purpose applications.

Available Accessories

Tampilkan Lebih Banyak