Detail AMS 5100 Z Gas Analyzer

The basics of the Oxygen Analyser AMS 5100 Z is the development project "AMS 5000". The Trace Oxygen Analyser AMS 5100 Z is designed for the measurement of the lowest Oxygen traces under presence of oxidable components for example in Air separation plants. For this application the Oxygen Analyser AMS 5100 Z is equipped with a catalytic inactive Zircondioxid sensor. Due to the compact design the Oxygen Analyser AMS 5100 Z is available in different housing types.

A long list of options allows to adopt the Oxygen Analyser AMS 5100 Z for almost every application. The software for remote control and remote service with access via the customer computer network further increase the flexibility of the Oxygen Analyser AMS 5100 Z in an application.

Technical data
Analyser
AMS 5100
Measuring principle
Catalytic inactive ZrO2 probe
Application
Gases Industries, Chemical Industries
Measuring range
0 … 25 Vol %
0 … 10 ppmv
Analogue signal port
(0) 4…20 mA or 0…10V, galvanically separated
Reproducability
+/- 2 % of the measuring value
Resolution
0,01 ppm – C(O2) – 0,01 %
depending on the O2 concentration
T90-Time
< 20 seconds
Display
2* 16 digit, illuminated LCD display
1. Line: display of concentration in ppm or Vol %
2. Line: Messages, device status, sensor parameters
Messages
1 System message (measuring value yes / no)
max. 1 message configurable as oxygen value,
calibration message, measuring value in the range, flow
Gas connection
inlet / outlet 3 / 6 mm ferrule pack
Gas sampling
built-in inlet / outlet valve
Sample flow
min. 20 Nl/h, max. 40 Nl/h
sample pressure (inlet)
min. 1,1 bar abs., max. 2 bar abs.
Digital communication
serial interface RS232
Ambient operating temp.
+ 5 °C up to + 60 °C
Power supply
110 … 230 VAC or 24 VDC
Protection / Housing / Dimensions
IP20 / electronics unit, 1/2 inch, 19 inch / 3HU
IP65 / wall mounting housing / ca. 327 x 270 x 326 mm (hxbxt)
IP54 / portable housing 42 TE / 3 HU
Weight
2,5 kg … 10 kg depending housing and built-in options
Options
Pneumatic pump
Electronic flow control
Software for remote control and service
Particle filter 2–7 µm
Pressure regulator
Tampilkan Lebih Banyak