Detail Rapidox 2100 OEM-SIL2 Gas Analyzer
The Rapidox 2100 OEM-SIL2 is a high-performance zirconia oxygen (O2) analyser, designed to meet the SIL2 requirements of IEC 61508/61511 under high demand use for safety integrity, as part of a Safety Instrumented System (SIS). The analyser design uses two microprocessors that continuously monitor the state of operation, quality of readings and the integrity of the digital & analogue outputs and the alarm relays. The likelihood of a fault being undetected is reduced by several orders of magnitude compared with a non-SIL2 equivalent product.
Product Specification
Type
Measurement
O2 Sensor Range
10E-20ppm to 30% Zirconia version. 10E-26 ppm extended range available on request
O2 Sensor Accuracy
±1% of the actual measured oxygen content OR 0.5ppm (whichever is the greater)
Ambient Operating Temperature
5-35°C 0-95% RH non condensing
Ambient Operating Pressure
800 to 1200mbar absolute
Max Sample Gas Pressure
Up to 10 bar gauge (200bar burst pressure)
Max Sensor Temperature
650°C
O2 Sensor Response
4 seconds for a T90 step change @1L per min flow
Life Expectancy
>17,000 hours
Sensor Cable
2m high temp as standard. Any length up to 25m available on request
Voltage
24V VDC +/-10%
Voltage Outputs
0-5V (0-10V on request) - continuously self-tested under SIL2
Current Outputs
4-20mA - continuously self-tested under SIL2
Digital Outputs
RS232 / RS485 & Modbus RTU - continuously self-tested under SIL2
Sample Connections
Nipple or Swagelok
Display
OLED display & keypad on enclosure version (not in the scope of SIL2)
Dimensions
70mm x 130mm x 105mm
Weight
<0.5kg in enclosure, OEM board 120g
Power / Warm-up Time
20W / 1-2 minutes at 20°C
Alarms
2 alarm relay circuits, fully user-configurable - self-tested every 10 minutes under SIL2
Calibration
Any 2 or 3 gases - Pre-calibrated Sensors available
UL Compliance
All paints, cables, wires, plastic materials etc are UL certified and marked
Enclosure Materials
Painted Aluminium
Din Rail Option
DIN Rail Mountable enclosure
Functional Safety
Meets the requirements of SIL2 under IEC 61508/61511